Assay types: End-point, Nephelometery, Multi-standards, Kinetic, Two-point, Enzymatic standard, Absorbency, Sample blank, Double wavelength. Factor counting automatically.
Wavelength range: 330-800nm; 340nm, 405nm, 510nm, 546nm, 578nm, 630nm fixed position. Two extra empty position for your selection. Wavelength range: ± 2nm, half wave band: ¡Ü8nm Light source: Halogen lamp, 6V/10W, ¡Ý3000 work hours + 4000 dormancy hours, auto dormancy function Flow cell: Ceramic frame with quartz window, 30¦Ìl flowing volume Temperature control: Room temperature, 25, 30, 37± 0.1º C Samples drawing volume: ¡Ü3000 Light absorbency range: -0.3000-3.0000ABS Resolution factor: ¡Ü0.001ABS(display), 0.00005ABS (internal computing) Carry over: 1% Repeatability, ¡Ü0.005ABS, CV¡Ü1%, Stability: ¡Ü0.005ABS/H Storage capacity: >100 test items, 1000 test records, 30 days Q. C. Memories, power cutoff protecting for ten years Man-machine interface: 5 inch display Man-machine conversation touch screen Display: Item parameters, processed data, reaction curve, Q. C. Records Communication: RS232C bi-directional Communication I/O, optional MODEM Report printer: Built-in 57mm high-speed thermal matrix printer. Environment: Ambient temperature 5-40º C Relative humidity ¡Ü90%, continuous working on ¡Ý12 hours External dimensions: 500(L)X340(W)X128(H)mm Power supply: AC150-260V, 50Hz, ¡Ü100VA Weight: 7Kg. |